X-RAYS IN THE HEALING ARTS

333-106-0050
General Requirements: Personnel Monitoring

All individuals who are associated with the operation of an X-ray system are subject to the requirements of OAR 333-120-0100 and 333-120-0210. In addition:

(1) When protective clothing or devices are worn on portions of the body and a personnel monitoring device(s) is required, at least one such monitoring device shall be utilized as follows:

(a) When an apron is worn, the monitoring device shall be worn at the collar outside of the apron;

(b) The dose to the whole body based on the maximum dose attributed to the most critical organ (which are the gonads, the blood-forming organs, head and trunk or lens of the eye), shall be recorded in the reports required by OAR 333-120-0650(3). If more than one device is used and a record is made of the data, each dose shall be identified with the area where the device was worn on the body.

(2) Exposure of a personnel monitoring device to deceptively indicate a dose delivered to an individual is prohibited.

Statutory/Other Authority: ORS 453.605 - 453.755
Statutes/Other Implemented: ORS 453.605 - 453.755